Uploaded by User37890

15465934178488

advertisement
Name of Product : Innova IGS 5
Brand : GE
Manufacture : France
Innova IGS 5
Gantry
C-arm Configuration
C-arm depth
C-arm movement
C-arm anti collision
Source Image Distance
Programmable gantry positions
Tube
Focal spot size
Anode Heat capacity
Note
Floor mounted L-arm and offset C-arm
The offset arm throat depth of 107 cm (42 in) with L-arm at 0° (Head
end), provides femoral coverage on most patients without rotating the Larm
Motorized
InnovaSense advanced patient contouring technology and press sensor
on tube cover also detector cover
89 cm to 119 cm
Yes
0 .3, 0 .6 and 1 .0 mm
Maximum casing heat storage:
2.74 MJ (3.7 MHU)
5.14 MJ (6.9 MHU)
Anode steady state heat dissipation:
6.72 kW
Continuous casing heat dissipation:
3.2 kW
Closed-loop remote water chiller
Tube Cooling System
Generator
Power
Maximum kVp available
Recording mode
Recording mA rating
Fluoroscopy mode
Fluoroscopy mA rating
Fluoroscopy timer
100 kW
125kVp
50 to 125 kVp
1 - 1000mA
60 to 120 kVp
1 - 140 mA
Yes
Flat Panel, Solid-state Detector
Detector Type :
Detector Size* :
Amorphous silicon photodiode array on a continuoussubstrate, singlepiece panel with no inherent seams.
A) 20.5 cm x 20.5 cm ;
B) 31 cm x 31 cm;
C) 41 cm x 41 cm.
* Hanya dapat dipilih 1 ukuran detector saja
DQE
80% for 20.5 Detector size ; 84% for 31 cm and 41 cm detector size
Image Display Monitor
3 (two) B&W 48 cm (19 in) LCD monitors for in-room display of live and reference images.
3 (two) ≥48 cm (19 in) in-room Color LCD monitors for AW and hemodynamic
Positioner indicators are integrated into in-room display for streamlined monitor configuration and reduced eye
distraction.
StopWatch in-room display is activated from Innova Central
User Interface
Dedicated keypad for convenient control of commonly used review functions provides an image shuttle knob to
control playback and one-touch access to review functions such as play/pause, previous/next image or run, zoom
and roam, edge enhancement adjustment , and store/recall reference images
Keyboard for patient data entry
Innova OneTouchQA at table-side QCA
Wireless remote for in-lab control of commonly used image playback and processing functions.
Auto-positioning
Touch Screen Interface beside table
- Acquisition preference settings
- Fluoro levels
- Field-of-view
- Fluoro landscape
- Fluoro timer reset
- Fluoro Loop and Store
Quantitative Analysis
Includes distance measurement, stenosis analysis and left ventricle analysis. Distance measurement and stenosis
analysis are also available in-room on the Innova Central Touch screen through the Innova OneTouchQA
Image Acquisition
Dynamic acquisition
DSA
Fluoroscopy
7.5 ; 15 and 30 fps
0 .5 - 7 .5 fps
3.75, 7.5, 15, 30 fps
Dose management and display
- Dose Monitoring
- Dose IQ customization
- Visualization of low contrast objects
- Dose Structured Reporting
- Dose Map - real time patient dose distribution map
Rumah sakit hanya dapat memilih
salah satu dari ukuran detector ini (A
atau B atau C) dan menuliskan di
dalam kolom komentar yang tersedia
Advanced applications
StentViz, Zoom and Enhance image of the stent, with automatic guidewire subtraction
StentVesselViz, fusion of enhanced stent with injected vessel
InnovaSpin, Rotational Angiography
InnovaChase, for unsubtracted visualization of a run off
Innova 3D, 3D imaging for internal body structures and interventional devices in one image
Innova CT, Crossectional imaging
Valve Assist, 3D guidance for structural heart procedures
In-room Browser, Users can review a thumbnail display of acquired sequences directly on the in-room
monitor for interactive tableside selection and review
Fluorostore, stores, displays, and plays loops of the last 450 fluoro images at the push of a button for
streamlined image review, helping to avoid extra images and exposure.
Power Distribution
Nominal Voltage :
Maximum momentary power demand:
Table
Omega V Table
Maximum table load
Length
Longitudinal travel
Lateral travel
Width
Horizontal Float Movement
Height range
380 to 480 Volts AC, three phase 50 or 60 Hz without neutral.
150 kVA
304 kg
333 cm
Up to 170 cm
± 14 cm
46 cm
8 - way
78 cm - 108 cm
Download